Summary | 0000380: There is the problem on the correlator | ||||||||
Description |
There is still the correlator problem which causes the different problem with different settings: In settings of 8 bands,4 polars,128ch,BW=10MHz(recirculation factor 2): The cross correlations of 0-7YDY have no Ampls in band2,6. In settings of 8 bands,2 polars,256ch,BW=10MHz(recirculation factor 2): All cross correlations of *XDX,*YDY have no Ampls in band5 and no TPs of DX,DY in band5. In settings of 8 bands,4 polars,64ch,BW=20MHz: All cross correlations of *YDY have no Ampls in band5 and no TP of DY in band5. (Sjouke and Albert: Above informations may be helpful for you to locate the problem in the correlator. For 4 polars data only know the problem on XX,YY but not on XY,YX.) |
||||||||
Additional Information | |||||||||
nr subbands | 1 | ||||||||
nr polarisations | |||||||||
nr channels | |||||||||
bandwidth | |||||||||
multi-subarray | no | ||||||||
recirculation | yes | ||||||||
Attached Files | |||||||||
|